OpStar İnfusion Pump WIT-601A
- Fast value setting with the keypad
- Ability to record sets, quick selection
- Sorting treatment with gradual mode
-Ability to send automatic bolus
Speed that can run in 6 different smarter modes; time, patient records; Drug Library Programmable sequential infusion, ability to reduce or increase the speed when running pigyback |
More useful: open system can be used with all universal sets Digital keyboard and menu whose values can easily be adjusted |
More secure double process lock mechanism; antireverse feature | More comfortable: Liquid heater (Optional) |
FEATURES | |
INFUSION SET OPTIONS: CAN ALSO BE USED SET-DEFINED IF DESIRED 10d/ml, 15d/ml, 20d/ml, 60d/ml(3.4mm 4.5mm) |
INFUSION MODES: SPEED MODE; TRICKLE MODE; TIME MODE; SEQUENTIAL PROGRAMMING MODE; WEIGHT (DOSE MODE); DRUG LIBRARY MODE |
STANDARD DEVIATION ERROR VALUES: MAXIMUM ± 5% | FLOW RATE: 0.1 ~ 1200ml/hour (20 d/ml IV set)/1 ~ 150 ml/hour (60 d/ml IV set) |
ALARMS: Air bubble; door open; completed; Stopped; Time-out; Empty, leakage, battery damaged, motor control, unidentified infusion set, drop sensor error, fuser error, upper track congestion, lower track congestion, connection error |
CONGESTION PRESSURE: Upper Pressure Standard; Lower Pressure Adjustable Value High 70-90KPA; MEDIUM 55-70KPA; LOW 40-55KPA |
DROP MODE SPEED: 1 ~ 400 drop/min (20 drop/ml IV set) 1 ~ 150 drop/min. (60DROP/ml IV set) | PROCESSOR: DUAL CPU |
ADJUSTABLE TOTAL DOSE: 1-9999ml | LOCK: AUTOMATIC LOCKING AGAINST UNAUTHORISED INTERVENTION |
LANGUAGE: ENGLISH RUSSIAN TURKISH | PROGRAMMING: Infusion treatment, eights-stage, programmable |
KVO 1ml/hour | DRUG LIBRARY: The user can define medication up to 100 pieces |
Purge 600 ml/HR | PATIENT RECORDS: Ability to transfer 1500 patient records to computer through LAN |
BOLUS: Opening value 600 ml/hour, total 2 ML, values can be increased if desired | PIGGYBACK: can be increased while speed is running without stopping the device. |
AIR BUBBLE DETECTION: Ultrasonic detector can set values. 50 ~ 800μL | FLUID HEATER: Optional |
Battery: 7 hours of duration, to full charge in 12 hours, CHARGING APPARATUS SUTITABLE FOR AMBULANCE DC 9.6 V, 2000mAh; | DIMENSIONS: 130 mm × 120 mm × 195 mm Weight: 2 KG |
OPERATIONAL CONDITIONS: TEMPERATURE 5 °C ~ 40 °C HUMIDITY 20% ~ 90% PRESSURE: 70 ~ 106KPa |
APPLICABILITY: Any fluids and medications in micro or rapid intravenous ELEKTİRKSEL SAFETY CLASS: Class I with inner power, Ipx4, CF |
POWER SOURCE: AC: 100V ~ 240V, 50 Hz/60 Hz; Power dissipation: 25VA |