RESCUE LIFE 7 Defibrillator

RESCUE LIFE 7 Defibrillator

GENERAL REQUIREMENTS
Dimensions 369 x 240 x 340 mm
Weight 5.5 Kg with battery
Power Supply AC 100-240 VAC, 50/60 Hz inverter 12V car adapter (optional)
Rechargeable 9.3 V-3.2 Al lithium-ion battery
Battery Operating Time 5 Hours
Battery Charge Time Max. 3 hours
Capacity 150 shocks in 230J, when battery fully charged 
Software Multilingual, voice prompted and in text
Memory 4 Gb internal memory, 300 hours of ECG data, reviewing events and usage
Review and printing of the data base Review and printing of archived events and
DEFIBRILLATION
Waveform Biphasic Truncated Exponential (BTE)
HiCAP technology (large storage capacitor)
Impedance Corrected, range 25-200 Ohm
Charging Time 8 seconds for 230J when battery fully charged
MANUAL MODE
Energy range 1-230K (1J'lik artışlar halinde 1-10 arası;
10J'lik adımlar halinde 10-23 J arası)
Commands Multi-function rotary setting button. For hands free defibrillation,
charging and shock buttons directly accessible on the front face of the panel
Pedals Reusable adult and child defibrillation pedals witj charge/shock commands
Disposable defibrillation pads and internal defibrillation pads (optional)
Operation Mode ECG «R» wave synch or asynchronous modes
Indicators Battery and main led indicators
Clear and visible backlight color buttons
AED MODU
Energy 200J'de sabitlenmiş enerji
Protocol Audio and text prompts as per ERC 2015 CPR Guide
Shockable Rhythms VF with amplitude > 0.15 mV, and VT with a rhythm of >150 min/pulse
Sensitivity SensitivityShockable Rhythm - VF > 95%
Shockable Rhythm - VT > 75%
SCREEN
Type High-contrast TFT color LCD 7" (800 X 480 pixels)
Information Operational information, ECG traces, SpO2 value, NIBP values,
parameters, pulse indicator, filters, alarms, and battery status
Display mode 3, 6, 12 ECG cable selection
ECG DISPLAY
Bandwidth 0.5-120 Hz (-3 db), when filters turned off
CMRR > 95 db
Input Impedance > 20 Mohm
Input 5-lead patient cable (I, II, III, aVR, aVL, aVF, V)-standard
10-lead patient cable (I, II, III, aVR, aVL, aVF, V1 ... V6) - optional
Gain with patient cable; 2.5, 5, 10, 20, 40 mm/mV. Automatic, with pads
Speed 5, 10, 25, 50 mm/sec
Filters 550/60 Hz, EMG Filter, start
Display Mode: bandwidth 0.6 - 40 Hz (-3 db)
Diagnostic Mode: bandwidth 0.05 - 120 Hz (-3 db)
Heart Rate range 20-300 BPM (~ 2%), digitally displayed on the screen
Alarm Max. adjustable heart rate, 250 beats/min; min. 20 BPM.
SPO2 (Optional)
SpO2 range 0 - 100*%
Heart Rate range 30 - 250 bpm
Accuracy For adults  70%- 100% ~ 2% with the finger sensor 
Alarm Min. adjustable value 50%
NIBP (OPTIONAL)
Technical Oscillometric
NIBP Accuracy meets ANSI/AAMI SP10-2002, EN 1060-4 standards
Patient Application Adult/Pediatric/Neonatal
Systolic Range Adult: 40-260mmHG, Pediatric: 40-160
mmHg, newborn 40-130 mmHG
Range MAP Adults: 26-220 mmHG, Pediatric: 26-133
mmHg, newborn 26-110 mmHG
Diastolic Range Adult: 20-200 mmHG, Pediatric: 20-120
mmHg, newborn 20-100 mmHG
PR Range 30-220 BPM
PR Accuracy +/-2% or +/3 bpm
Operation Modes Manual, Automatic, Long-Term Automatic, Stat, Service
Automatic Interval Periods 1, 2, 3, 4, 5, 10, 15, 30, 60 and 90 minutes
PRINTER
Type Thermal 3-channel integrated printer to print documents
 regarding ECG traces and events including heart rate/SpO2/NIBP values
Paper Speed 5, 10, 25, 50 mm/sec
Paper Width 58 mm
Opertion Mode Manual, automatic (a 10-second record before and after Shock)
EXTERNAL PACEMAKER (OPTIONAL)
Type Rectangular wave
Operation Mode Fixed upon request
Pulse rate From 30 pulses/min to 250 pulses/min, adjustable in steps of 5 pulses/min
Response time 22.5 ms
Pulse current from 0 to 150 mA, can be set in steps of 5mA
Amplitude Max 150 V
STANDARDS AND SECURITY
Standard IEC 60601-2-4; IEC 60601-1; IEC 60601-1-2
CE Mark EEC 93/42 Medical Devices, Class IIb